Página 1

Mobile IV Stand — instructions for use
Model: IVSTND-260 | English | Rev 1.2 | 25 October 2026
Read these instructions before assembly and use. Keep them with the stand. This is a non-sterile, reusable
medical device for use by trained healthcare staff indoors.


Intended purpose
IVSTND-260 supports compatible infusion bags or bottles on four hooks and organises light clinical supplies
on a removable tray and lower basket. It also has supports for a waste bin and a compatible 1 L round sharps
container. Its contribution is mechanical support and access to supplies during care.
The stand does not pump, measure, administer or monitor an infusion and does not contact the fluid pathway.
It does not support or transport the patient.


Operating limits

Feature                                Limit or specification

Maximum combined load                  10 kg across the hooks, tray and basket

All four hooks combined                4 kg total, distributed evenly

Removable tray                         2 kg

Lower basket                           4 kg

                                       Minimum 1,200 mm · normal 1,750 mm · maximum 2,000 mm
Height adjustment
                                       Measured from the floor to the top of the hooks

Base                                   Five wheels, two fitted with brakes

Sharps container                       Compatible round 1 L container; follow its own instructions
The hook, tray and basket ratings may be used together within the 10 kg combined limit. Do not use
internal test loads as operating ratings.


Do not use the stand for
Sitting, standing, lifting, walking support or patient transport; mounting pumps or other equipment on the tube;
outdoor use; stairs; uneven or sloping floors. MRI suitability has not been established.
Stop use if the stand is unstable or a joint, height lock, wheel, brake or attachment is defective. Residual risks
include tipping or falling loads, unintended movement, pinching at clamps, snagged tubing, sharps injury and
cross-contamination from inadequate cleaning and disinfection. Keep within all load limits, use the brakes,
secure loads, keep hands and lines clear, and follow the sharps-container and cleaning instructions.




                                                                                     ECI-IFU-STND-260 | Rev 1.2 | Page 1

Página 2

Assembly
Follow the sequence below using the matching supplied parts and hardware. Do not substitute fasteners.

1 Prepare the telescopic pole




                                               Remove the hexagonal screw from the bottom of the pole.
                                               Align the hand-tightening screw with the hole at the
                                               telescopic connection and tighten it to secure the pole. No
                                               tools required.




2 Fit the tray support



                                               Slide the tray support onto the pole from the bottom and
                                               position it at the required height. Secure it with the black T-
                                               shaped screw supplied for the support. No tools required.



3 Attach the pole to the base




                                               Insert the plastic cover into the central hole of the base.
                                               Insert the pole and secure it from underneath with the M10
                                               × 35 hexagonal combination screw. Use the 16 mm open-
                                               ended wrench.




4 Fit the lower mounting plate



                                               Remove the clamp from the lower mounting-plate assembly.
                                               Position the plate and clamp around the pole as shown and
                                               secure them with two M6 × 65 socket-head screws. Use the
                                               5 mm hex key.




                                                                                 ECI-IFU-STND-260 | Rev 1.2 | Page 2

Página 3

Assembly continued
5 Fit the basket




                                                Align the basket with the threaded holes on the back of the
                                                mounting plate. Secure it with two M4 × 12 socket-head
                                                screws. Use the 3 mm hex key.




6 Fit the waste bin




                                                Align the waste bin with the two hooks on the mounting
                                                plate. Lower it onto the hooks and press down firmly. No
                                                tools required.




7 Fit the sharps-container holder




                                                Position the holder beneath the tray support, aligning the
                                                two fixing holes shown. Secure it with two M4 × 10
                                                countersunk hex-socket screws. Use the 2.5 mm hex key.




8 Fit the removable tray




                                                Lower the stainless-steel tray into the tray support. Check
                                                that it is seated securely before loading. No tools required.




After assembly, check every joint and attachment, both brakes and the height lock before placing the stand in
service.




                                                                                 ECI-IFU-STND-260 | Rev 1.2 | Page 3

Página 4

Before each use
• Check that the stand is clean, the identification is readable and the surfaces are intact.
• Check all joints, clamps and accessories for secure attachment. Check that the tray is seated properly.
• Check that all wheels turn freely and both brakes operate. Apply both brakes and check that the stand stays
stationary.
• Support the upper tube before loosening the height lock. Adjust within 1200–2000 mm, retighten the lock
and check that the tube does not slip before loading.
• Distribute the hook loads evenly and keep within all individual and combined load limits.


Stationary use and movement
Apply both wheel brakes during stationary use. Keep fingers clear of sliding and locking parts. Keep the
infusion tubing clear of the wheels and any parts that could trap or pull it.
Before moving the stand, secure the containers and retract the sharps-container holder beneath the tray.
Release the brakes and push slowly, keeping the load stable. Approach small door thresholds cautiously; stop
and obtain assistance if the stand is unstable. Do not force the loaded stand over an obstruction.


Using the sharps-container holder
Pull out the holder for use and return it beneath the tray for movement. Use the compatible round 1 L sharps
container. Follow the container manufacturer’s fill limit, closure and disposal instructions. Do not place sharps
in the ordinary waste bin. The holder does not replace the container’s protective features or instructions.


Cleaning and disinfection
Remove supplies and waste safely. Remove soiling with a soft cloth and mild detergent in water. Wipe off
detergent with a clean damp cloth and allow the surface to air dry. Do not mix chemicals, immerse the stand,
pressure-wash it or use abrasive cleaners.
Disinfect with ready-to-use hospital surface-disinfectant wipes based on 70% isopropyl alcohol. Keep the
surface visibly wet for the full contact time stated on the wipe label, re-wetting if needed. Do not use other
disinfectant types.


Maintenance, service life and warranty
Specific scheduled maintenance: none. Carry out the before-use checks in this manual. Stop using a stand
with loose joints, a slipping lock, faulty brakes, damaged parts or surfaces that cannot be cleaned. Qualified
personnel must repair it and verify safe function before reuse.
Expected service life: at least five years under normal use, subject to the condition checks in this manual.
Warranty: two years. The expected service life is not an extension of the warranty.




                                                                                    ECI-IFU-STND-260 | Rev 1.2 | Page 4

Página 5

End of use
Decontaminate the stand before disposal. Handle sharps and clinical waste separately under the applicable
local rules and the container instructions. Dispose of the stand’s materials in accordance with local
requirements.


Identification and reporting
Model: IVSTND-260. GTIN: 6980438430027. Identify an individual unit by its serial number and lot number on
the product label. Quote the model, serial number and lot when contacting EverCare about a defect or
incident. The 14-digit UDI-DI representation is 06980438430027. The house symbol on the label means
"Indoor use only".
Report any serious incident associated with the device to the manufacturer and to the competent authority of
the Member State where the user or patient is established.


Manufacturer
EVERCARE INNOVATIONS (SHENZHEN) CO., LTD.
Unit 102, Building J, No. 3 Xiang Yuan Pai Road, Pingxi Community, Pingdi Subdistrict, Longgang District,
Shenzhen, Guangdong 518000, China
Telephone: +86 136 3167 8692
Email: service@evercare-innovations.com
Website: www.evercare-innovations.com


European authorised representative
MDSS GmbH
Schiffgraben 41, 30175 Hannover, Germany
Document ECI-IFU-STND-260 | Rev 1.2 | 25 October 2026




                                                                                ECI-IFU-STND-260 | Rev 1.2 | Page 5

Página 6

Identification and document control
Mobile IV Stand
Medical device | Non-sterile | Reusable | Professional healthcare use

Model / catalogue number       UDI-DI (GTIN, 14-digit representation)

IVSTND-260                     06980438430027
The UDI-DI identifies the product model, not an individual unit. Read the unit’s serial number and lot number
from its product label. Match the model and unit identification on the device, packaging and accompanying
records before use. Do not substitute the manufacturer SRN for a product UDI.

Manufacturer identification
EVERCARE INNOVATIONS (SHENZHEN) CO., LTD.
Unit 102, Building J, No. 3 Xiang Yuan Pai Road, Pingxi Community, Pingdi Subdistrict, Longgang District,
Shenzhen, Guangdong 518000, China
Manufacturer SRN: CN-MF-000060301
Email: service@evercare-innovations.com | Telephone: +86 136 3167 8692
Website: www.evercare-innovations.com

EU authorised representative
MDSS GmbH
Schiffgraben 41, 30175 Hannover, Germany

Revision identification
ECI-IFU-STND-260 | Rev 1.2 | English | 25 October 2026
Changes since released Rev 1.0: full manufacturer address and identification page (manufacturer SRN,
model and GTIN, authorised representative and CE marking); residual risks; and cleaning and disinfection
instructions. The indoor-use-only label symbol is explained. Revision 1.1 was an unreleased draft.




                                                                                 ECI-IFU-STND-260 | Rev 1.2 | Page 6